--- 美國(guó)SPEX CLP項(xiàng)目標(biāo)液 ---
ICP-MS干擾測(cè)試標(biāo)液B1
ICP-MS Analytes B1
元素:20mg/L: Co, Cr, Cu, Mn, Ni, V
10mg/L: As, Cd, Se, Zn
5mg/L: Ag
基體:2% HNO3
體積:125mL
編號(hào):CL-INT-B1
用途:Claritas PPT 級(jí),電感耦合等離子體質(zhì)譜儀(ICP-MS)分析用元素標(biāo)液。EPA合約實(shí)驗(yàn)室CLP項(xiàng)目標(biāo)液。
證書:每份標(biāo)液都附有一份內(nèi)容詳細(xì)的標(biāo)準(zhǔn)物質(zhì)證書,證書報(bào)告標(biāo)液中多至68種痕量雜質(zhì)元素含量實(shí)際測(cè)量結(jié)果。
溯源:美國(guó)NIST SRM (Traceable by NIST SRM)
**應(yīng)用參考 · INSTRUCTION**
每次測(cè)試開(kāi)始和結(jié)束驗(yàn)證干擾和背景校正因子,另外每20個(gè)樣品必需進(jìn)行檢測(cè)。標(biāo)液CL-INT-A1,稀釋20倍,為溶液A,標(biāo)液CL-INT-A1稀釋20倍再加標(biāo)液CL-INT-B1稀釋100倍后混合,為溶液AB。溶液A和溶液AB用于見(jiàn)的干擾檢測(cè)。一旦兩種溶液配制好,需連續(xù)分析,從溶液A開(kāi)始。
CLP-M/6020/SW-846 Standards for ICP-MS
For verification of inter-element and background correction factors at the beginning and the end of each analysis run. In addition, a verification must be done after every 20th sample. Two solutions are required for the most common interference check: Solution A, the interferents alone (CL-INT-A1) and Solution AB, a combination of interferents (CL-INT-A1) and analytes (CL-INT-B1). Solution A is prepared by diluting CL-INT-A1 20-fold. Solution AB is prepared by diluting CL-INT-A1 20-fold and CL-INT-B1 100-fold; for example, 5 mL of CL-INT-A1 and 1 mL of CL-INT-B1 into a 100 mL volumetric flask, brought to volume with a matrix blank . Once prepared, the solutions should be analyzed consecutively, starting with Solution A.